TY - JOUR
T1 - Comparison of image quality between Deep learning image reconstruction and Iterative reconstruction technique for CT Brain- a pilot study
AU - Chandran M, Obhuli
AU - Pendem, Saikiran
AU - Priya, P. S.
AU - Chacko, Cijo
AU - Priyanka, null
AU - Kadavigere, Rajagopal
N1 - Publisher Copyright:
Copyright: © 2024 Chandran M O et al.
PY - 2024
Y1 - 2024
N2 - Background: Non-contrast Computed Tomography (NCCT) plays a pivotal role in assessing central nervous system disorders and is a crucial diagnostic method. Iterative reconstruction (IR) methods have enhanced image quality (IQ) but may result in a blotchy appearance and decreased resolution for subtle contrasts. The deep-learning image reconstruction (DLIR) algorithm, which integrates a convolutional neural network (CNN) into the reconstruction process, generates high-quality images with minimal noise. Hence, the objective of this study was to assess the IQ of the Precise Image (DLIR) and the IR technique (iDose 4) for the NCCT brain. Methods: This is a prospective study. Thirty patients who underwent NCCT brain were included. The images were reconstructed using DLIR-standard and iDose 4. Qualitative IQ analysis parameters, such as overall image quality (OQ), subjective image noise (SIN), and artifacts, were measured. Quantitative IQ analysis parameters such as Computed Tomography (CT) attenuation (HU), image noise (IN), posterior fossa index (PFI), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the basal ganglia (BG) and centrum-semiovale (CSO) were measured. Paired t-tests were performed for qualitative and quantitative IQ analyses between the iDose 4 and DLIR-standard. Kappa statistics were used to assess inter-observer agreement for qualitative analysis. Results: Quantitative IQ analysis showed significant differences (p<0.05) in IN, SNR, and CNR between the iDose 4 and DLIR-standard at the BG and CSO levels. IN was reduced (41.8-47.6%), SNR (65-82%), and CNR (68-78.8%) were increased with DLIR-standard. PFI was reduced (27.08%) the DLIR-standard. Qualitative IQ analysis showed significant differences (p<0.05) in OQ, SIN, and artifacts between the DLIR standard and iDose 4. The DLIR standard showed higher qualitative IQ scores than the iDose 4. Conclusion: DLIR standard yielded superior quantitative and qualitative IQ compared to the IR technique (iDose4). The DLIR-standard significantly reduced the IN and artifacts compared to iDose 4 in the NCCT brain.
AB - Background: Non-contrast Computed Tomography (NCCT) plays a pivotal role in assessing central nervous system disorders and is a crucial diagnostic method. Iterative reconstruction (IR) methods have enhanced image quality (IQ) but may result in a blotchy appearance and decreased resolution for subtle contrasts. The deep-learning image reconstruction (DLIR) algorithm, which integrates a convolutional neural network (CNN) into the reconstruction process, generates high-quality images with minimal noise. Hence, the objective of this study was to assess the IQ of the Precise Image (DLIR) and the IR technique (iDose 4) for the NCCT brain. Methods: This is a prospective study. Thirty patients who underwent NCCT brain were included. The images were reconstructed using DLIR-standard and iDose 4. Qualitative IQ analysis parameters, such as overall image quality (OQ), subjective image noise (SIN), and artifacts, were measured. Quantitative IQ analysis parameters such as Computed Tomography (CT) attenuation (HU), image noise (IN), posterior fossa index (PFI), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the basal ganglia (BG) and centrum-semiovale (CSO) were measured. Paired t-tests were performed for qualitative and quantitative IQ analyses between the iDose 4 and DLIR-standard. Kappa statistics were used to assess inter-observer agreement for qualitative analysis. Results: Quantitative IQ analysis showed significant differences (p<0.05) in IN, SNR, and CNR between the iDose 4 and DLIR-standard at the BG and CSO levels. IN was reduced (41.8-47.6%), SNR (65-82%), and CNR (68-78.8%) were increased with DLIR-standard. PFI was reduced (27.08%) the DLIR-standard. Qualitative IQ analysis showed significant differences (p<0.05) in OQ, SIN, and artifacts between the DLIR standard and iDose 4. The DLIR standard showed higher qualitative IQ scores than the iDose 4. Conclusion: DLIR standard yielded superior quantitative and qualitative IQ compared to the IR technique (iDose4). The DLIR-standard significantly reduced the IN and artifacts compared to iDose 4 in the NCCT brain.
UR - https://www.scopus.com/pages/publications/85197479400
UR - https://www.scopus.com/inward/citedby.url?scp=85197479400&partnerID=8YFLogxK
U2 - 10.12688/f1000research.150773.1
DO - 10.12688/f1000research.150773.1
M3 - Article
C2 - 38962692
AN - SCOPUS:85197479400
SN - 2046-1402
VL - 13
JO - F1000Research
JF - F1000Research
M1 - 691
ER -