Yet another cause for blocked sidestream capnogram - Beware of the non-threaded cap mount in heat and moisture exchangers

Goneppanavar Umesh, Kaur Jasvinder, Nanda Shetty

Research output: Contribution to journalArticlepeer-review

Abstract

Heat and moisture exchangers (HME) are commonly used during general anaesthesia and intensive care of patients on mechanical ventilators. Some of the HME manufacturers provide HMEs with a Luer lock fitting for connecting side stream CO2 monitoring line, Luer lock cap, and a non-threaded cap mount. However, HMEs from different manufacturers and HMEs meant for use in children and for adults from the same manufacturer vary in the presence/absence of non-threaded cap mount. This can create confusion to the clinicians and can result in inadvertent connection of the CO2 monitoring line to the non-threaded cap mount resulting in blocked CO2 monitoring line and leak in the circuit. We caution all the anaesthesiologists and intensivists regarding this possibility while using HMEs from different manufacturers.

Original languageEnglish
Pages (from-to)207-208
Number of pages2
JournalJournal of Clinical Monitoring and Computing
Volume23
Issue number4
DOIs
Publication statusPublished - 09-06-2009
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Medicine(all)
  • Health Informatics
  • Critical Care and Intensive Care Medicine
  • Anesthesiology and Pain Medicine

Fingerprint

Dive into the research topics of 'Yet another cause for blocked sidestream capnogram - Beware of the non-threaded cap mount in heat and moisture exchangers'. Together they form a unique fingerprint.

Cite this